proudly presents the Turner Designs

10-AU-005-CE Field Fluorometer
Application Configuration Guide

Many of the links on this page go to the Turner Designs web site. Use the [BACK] button on your browser after activating these links to return to this page.

10-AU Fluorometer/Fluorimeter

Instrument

Optical Configuration


Application

Instrument
P/N

PMT
(installed)

*Optical Kit P/N

Excitation Filter P/N & Description

Emission Filter P/N & Description

Reference Filter P/N & Description

Lamp P/N & Description

Attenuator Plate

Solid
Secondary
Standard

Chlorophyll a, Extractive and In-vivo (Acidification method)

10AU-072

Red
185-870nm

10-037R

10-050R
340-500nm

10-051R
>665nm

10-032
1 ND Square

10-045
Daylight White

Not Required

10-AU-904

Chlorophyll a, Extractive
(5Ex.436/Em.680nm)
(Non-acidification method)

10AU-072

Red
185-870nm

10-040R

10-113
436nm

10-115
680nm

10-032
1 ND Square

10-089
Blue Mercury Vapor

Not Required

10-AU-904

Chlorophyll a, In-Vivo
(Water with high dissolved organic matter(DOM) and/or chlorophyll b )

10AU-072

Red
185-870nm

10-096R

10-050R
340-500nm

10-115
680nm

10-032
1 ND Square

10-089
Blue Mercury Vapor

Not Required

10-AU-904

Fluorescein (5Ex.490/Em.580nm)

10AU-073

Standard
300-650nm

10-086R

10-105
486nm

10-109R-C
510-700nm

10-063
>485nm

10-089
Blue Mercury Vapor

10-318R2
(if needed)

Rhodamine WT3
(5Ex.550nm/Em.580nm)


NOTE!
For Standard PMT Instruments, the
10-058R Emission Filter should not be used!

10AU-073

Standard
300-650nm

10-041R

10-103
550nm

10-052R6
>570nm (near PMT) &
10-058R4
325-700nm
(near sample)
for Red PMT instruments only

10-053
>535

10-046
Clear Quartz1

Not Required

10-AU-904

Oil, Short Wavelength
(5Ex.254nm/Em.300-400nm)
USE A QUARTZ CUVETTE!

10AU-073

Standard
300-650nm

10-301R

10-038
254nm

10-069R
300-400nm

10-300
Soft Glass
>300nm

10-046
Clear Quartz1

10-327R2 or
10-318R2
(if needed)

Oil, Long Wavelength
(5Ex.360nm/Em.410-600nm)

10AU-073

Standard
300-650nm

10-302R

10-069R
300-400nm

10-110R-C
410-600nm

10-300
Soft Glass
>300nm

10-049
Near U.V.

10-327R2 or
10-318R2
(if needed)


*If Optical Kit P/N is ordered, kit will include excitation and emission filters plus two lamps and attenuator plates if applicable.

NOTES:

For detailed information about installing filters, refer to your Model 10-AU Fluorometer User's Manual.

1. WARNING! Do not let light from the clear quartz lamp fall on your eyes. Serious eye irritation will result, with a delay of about four hours between exposure and symptoms.

2. 10-327 Attenuator Plate. For highly fluorescent hydrocarbons, sometimes the fluorometer is too sensitive even with the sensitivity set to minimum. Try calibrating without an attenuator plate first. Then, try calibrating with the 10-318 Attenuator Plate installed(see next paragraph). Lastly, if the instrument is still too sensitive, you can reduce the sensitivity 75-fold by installing the 10-327 plate. For instructions, see the Filter and Light Source Replacement appendix in your user's manual.

10-318 Attenuator Plate. For highly fluorescent hydrocarbons with both the Short Wavelength and the Long Wavelength Oil Optical Kits and for some in vivo measurements with the Chlorophyll Optical Kit, sometimes the fluorometer is too sensitive even with the sensitivity set to minimum. Also, using the Rhodamine and Fluorescein Optical Kits, we have found that background fluorescence can be very high in natural systems; even with the sensitivity at minimum, readings may be over-scale or blank unduly high. In these cases, try calibrating without an attenuator plate first. Then, if necessary, you can reduce the sensitivity 5-fold by installing this plate. For instructions, see the Filter and Light Source Replacement appendix in your user's manual.

3. This filter combination applies equally well for Rhodamine B, Rhodamine WT, Sulpho Rhodamine B extra, and Pontacyl Brilliant Pink B.

4. Expect a significant temperature coefficient when using this emission filter combination; for best results, warm-up the instrument!

5. These excitation (Ex) and emission (Em) wavelengths are for the substances themselves, not the filters. Oils, for example, have a range of peak wavelengths, depending on the type of oil.

6. The 10-058R emission should only be used with red-sensitive PMT instruments or in environments with high algal levels. For instruments with standard PMTs, the10-052R is the only emission filter required for the Rhodamine optical kit.

Installation of Filters & Lamp